729/729FC
Users Manual
70
Table 8. Error Messages (cont.)
Error Message Explanation Potential Root Cause
Lost connection to HART
transmitter, Task aborted
Running task aborted, due to lost connection
to HART transmitter.
Make sure the transmitter is connected and
powered. Try HART polling again. After
connected, rerun the task.
Lost connection to FC devices,
Task aborted
Running task aborted, due to lost connection
to Fluke Connect secondary unit.
Turn on the FC secondary device and make
sure FC is active. Redo the discovery
process. After connected, run the task
again.
Lost connection to RTD, Task
aborted
Running task aborted because RTD was
removed.
Make sure the RTD probe is connected and
run the task again.
Open circuit detected Open circuit detected while sourcing mA. Connect the circuit properly and try again.
Invalid Task Setting Task setting contains invalid items.
Verify zero/full-scale value defined in the
task are in the range of the selected
pressure source and try again.