EasyManua.ls Logo

Fluke 8062A - High Impedance DC Voltage

Fluke 8062A
124 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
8062A
Instruction Manual
2-14
Since average-responding meters have been in use for so long, you may have
accumulated test or reference data based on them. The conversion factors in
Figure 2-8 should help you convert between the two measurement methods.
2-15. High Impedance DC Voltage
Occasionally you may want to make dc voltage measurements in high
impedance circuitry where even the 10 M
input impedance for the normal
dc voltage function could load the circuit and cause significant errors. For
example, a 10 M
input impedance causes a .1% error when measuring the
voltage across the 10 k
leg of a 90 k
over 10 k
voltage divider. The
8062A offers a >1000 M
(typically >10,000 M
) input impedance dc
voltage function which greatly reduces this error.
Figure 2-9 describes how to select the high input impedance dc voltage
function (the ac voltage function does not operate in this mode). Notice that
all of the function switches must be out to select this function. Either the 2V
or the 200 mV ranges may be selected.
2000mA
A
A
COMMON
V S
V
S
200mA
200
200k
200µA
DC
AC
200mV
200
20mA
20
20k
2mA
2
2k
1000 DC
750 AC
M
REL
1000V DC
750V AC
MAX
2A MAX
500V MAX
!
!
Low (-)
High Impedance
DC Voltage (V)
High (+)
1. Select the 2V or
the 200 mV range.
2. Ensure all function
switches are out.
3. Connect the test leads as shown.
4. Heed the input overload limits
(Table 2-2) and connect the leads to
the circuit being measured.
5. Read the measured value on the display.
dy10f.eps
Figure 2-9. High Impedance DC Voltage

Table of Contents

Related product manuals