Maintenance
Performance Tests
4
4-13
4-14. Voltage Test
Use the following procedure to verify proper operation of the dc and ac
voltage functions.
1. Connect the UUT and the DMM Calibrator as shown in Figure 4-4.
2. For each step in Table 4-2, select the UUT function and range as
indicated. Program the DMM Calibrator for the specified input signal
and verify that the displayed UUT value is within the indicated limits.
Table 4-2. Voltage Test
Step Function Range
Input Signal
Display
Level Freq.
Limits
1
2
3
4
5
6
DC Voltage 200 mV
200 mV
2V
20V
200V
1000V
+ 190.00 mV
-190.00 mV
1.9000V
19.000V
190.00V
1000V
dc +189.90 to 190.11
-189.89 to -190.11
1.8989 to 1.9011
18.985 to 19.015
189.85 to 190.15
999.1 to 1000.9
7
8
AC Voltage 200 mV 100.00 mV rms 200 Hz
20 kHz
99.40 to 100.60
98.60 to 101.40
9
10
11
12
13
14
15
AC Voltage 2V 1.0000V rms
1.0000V rms
1.0000V rms
1.0000V rms
1.0000V rms
200.0 mV rms
200.0 mV rms
200 Hz
1 kHz
10 kHz
30 kHz
20 kHz
200 Hz
30 kHz
.9940 to 1.0060
.9930 to 1.0070
.9930 to 1.0070
.9860 to 1.0140
.9890 to 1.0110
.1978 to .2022
.7940 to .2060
16
17
18
AC Voltage 20V 10.000V rms
10.000V rms
10.000V rms
200 Hz
10 kHz
30 kHz
9.940 to 10.060
9.480 to 10.520
9.460 to 10.540
19
20
21
22
23
AC Voltage 200V
200V
200V
750V
750V
100.00V rms
100.00V rms
100.00V rms
750.0V rms
750.0V rms
200 Hz
10 kHz
30 kHz
100 Hz
1 kHz
99.40 to 100.60
94.80 to 105.20
94.60 to 105.40
734.0 to 766.0
734.0 to 766.0