Performance Tests
Performance Test 3
3-59
8. Connect the thermal Power Sensor RF input directly to the UUT front panel
Microwave Output connector. See Figure 3-16 for equipment connections.
Power Meter
hpn369.eps
Figure 3-16. Microwave Output Level Accuracy Tests, High Frequency Points
9. Set the UUT as follows:
Mode Leveled Sine
Signal Microwave Output
Frequency 100 kHz
Level +13 dBm
Output OPER
10. Configure the Power Meter to indicate readings in dBm. Enter the
measurement frequency (0.1 MHz) into the Power Meter to enable cal factor
(frequency) correction.