96000 Series
Service Manual
3-58
4. Set the UUT output to the next frequency listed in Table 3-24 for this test
amplitude.
5. Repeat step 3 and confirm that the measured output level is consistent with
the typical performance shown in Table 3-24.
6. When the test frequency is 100 kHz, record the measured level in dBm as P
1
for use later in this procedure.
7. Repeat steps 2 through 7 for the next test amplitude listed in Table 3-24 for
the relevant UUT Low Level Microwave Output option configuration, applying
the 5790A 2.2 mV range linearity error correction at levels below -40 dBm.
Table 3-24. Microwave Output Level Accuracy Test, Low Frequency Test Points
Low Level Microwave Output
Option?
Amplitude Frequency Typical Tolerance
All UUT configurations
+13 dBm
1 kHz ±0.5 dB
100 kHz (P
1
) ±0.5 dB
+3 dBm
1 kHz ±0.5 dB
100 kHz (P
1
) ±0.5 dB
Non-Low Level MW O/P Option
only
-4 dBm
1 kHz ±0.5 dB
100 kHz (P
1
) ±0.5 dB
Low Level Microwave Output
Option only
-7 dBm
1 kHz ±2.0 dB
100 kHz ±2.0 dB
-17 dBm
1 kHz ±2.0 dB
100 kHz ±2.0 dB
-27 dBm
1 kHz ±2.0 dB
100 kHz (P
1
) ±2.0 dB
-37 dBm
1 kHz ±2.0 dB
100 kHz (P
1
) ±2.0 dB
-47 dBm
1 kHz ±2.0 dB
100 kHz (P
1
) ±2.0 dB