Performance Tests
Performance Test 3
3-57
hpn393.eps
Figure 3-15. Microwave Output Level Accuracy Tests, Low Frequency Points Connections
2. Set the UUT as follows:
Mode Leveled Sine
Signal Microwave Output
Frequency 1 kHz
Level +13 dBm
Output OPER
3. Allow the ac measurement standard to make several measurements and the
reading to settle. Convert the settled reading from V rms to dBm using the
following formula:
dBm (50 )
×
=
001.050
10
2
10
V
Log
The result must be within the tolerance listed in Table 3-24.