Performance Tests
Performance Test 3
3-61
Table 3-25. Microwave Output Level Accuracy Test, High Frequency Test Points
Low Level Microwave
Output Option?
Amplitude Frequency Reading P
out
Typical Tolerance
All UUT configurations
+13 dBm
100 kHz P
2
Reference
10 MHz P
3
±0.5 dB
100 MHz P
3
±0.5 dB
1 GHz P
3
±0.5 dB
4.024 GHz P
3
±1.0 dB
10 GHz P
3
±1.0 dB
15 GHz P
3
±1.0 dB
20 GHz P
3
±1.0 dB
26.4 GHz P
3
±1.0 dB
+3 dBm
100 kHz P
2
Reference
10 MHz P
3
±0.5 dB
100 MHz P
3
±0.5 dB
1 GHz P
3
±0.5 dB
4.024 GHz P
3
±1.0 dB
10 GHz P
3
±1.0 dB
15 GHz P
3
±1.0 dB
20 GHz P
3
±1.0 dB
26.4 GHz P
3
±1.0 dB
Non-Low Level MW O/P
Option only
-4 dBm
100 kHz P
2
Reference
10 MHz P
3
±0.5 dB
100 MHz P
3
±0.5 dB
1 GHz P
3
±0.5 dB
4.024 GHz P
3
±1.0 dB
10 GHz P
3
±1.0 dB
15 GHz P
3
±1.0 dB
20 GHz P
3
±1.0 dB
26.4 GHz P
3
±1.0 dB
Low Level Microwave
Output Option only
-7 dBm
100 kHz P
2
Reference
10 MHz P
3
±2.0 dB
100 MHz P
3
±2.0 dB
1 GHz P
3
±2.0 dB
4.024 GHz P
3
±2.0 dB
10 GHz P
3
±2.0 dB
15 GHz P
3
±2.0 dB
20 GHz P
3
±2.0 dB
26.4 GHz P
3
±2.0 dB