EasyManua.ls Logo

Fluke 96000 Series - Page 109

Fluke 96000 Series
198 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Performance Tests
Performance Test 3
3-67
!
!
!
MHz
Frequency Counter
Front
Rear
10 MHz Output from
Frequency Standard
External
Reference
Frequency
Input
Instrument
hpn391.eps
Figure 3-18. Equipment Connections - Microwave Output Frequency Accuracy Test Part A
4. Set the UUT as follows:
Ref Frequency Input Disabled
Mode Leveled Sine
Signal Microwave Output
Ref Frequency Input Disabled
Frequency 10 kHz
Level +13 dBm
Output OPER
5. Set the frequency counter as follows:
Function Freq A
Slope Positive
Coupling AC
Input Impedance 50
Ω
Attenuation 1x
Trigger Mode Auto
6. Set the output frequency of the UUT to each of the frequencies listed in part
A of Table 3-28. At each frequency allow the frequency counter to take
several readings and settle; the settled reading must be within the tolerance
shown. At 1 GHz connect the UUT Microwave output to the counter channel
C input and select the counter Freq C function.
7. Set the UUT output to STBY. Remove all connections to the frequency
counter.

Table of Contents

Related product manuals