Performance Tests
Performance Test 3
3-73
Table 3-29. Harmonics Test, Microwave Output
Level Frequency
2
nd
Harmonic
Limit
3
rd
Harmonic
Limit
+16 dBm
20 kHz 40 kHz -60 dBc 60 kHz -60 dBc
2.74 MHz 5.48 MHz -60 dBc 8.22 MHz -60 dBc
5.5 MHz 11 MHz -60 dBc 16.5 MHz -60 dBc
11 MHz 22 MHz -60 dBc 33 MHz -60 dBc
22 MHz 44 MHz -60 dBc 66 MHz -60 dBc
50 MHz 100 MHz -60dBc 150 MHz -60dBc
88 MHz 176 MHz -60 dBc 264 MHz -60 dBc
125 MHz 250 MHz -60 dBc 375 MHz -60 dBc
250 MHz 250 MHz -60 dBc 750 MHz -60 dBc
354 MHz 708 MHz -60 dBc 1.062 GHz -60 dBc
500 MHz 1 GHz -60 dBc 1.5 GHz -60 dBc
1 GHz 2 GHz -60 dBc 3 GHz -60 dBc
1.25 GHz 2.5 GHz -55 dBc 3.75 GHz -55 dBc
1.4 GHz 2.8 GHz -55 dBc 4.2 GHz -55 dBc
1.8 GHz 2.8 GHz -55 dBc 5.4 GHz -55 dBc
2 GHz 4 GHz -55 dBc 8 GHz -55 dBc
2.7 GHz 5.4 GHz -55 dBc 8.1 GHz -55 dBc
5 GHz 10 GHz -55 dBc 15 GHz -55 dBc
9 GHz 18 GHz -55 dBc - -
10 GHz 20 GHz -55 dBc - -
12 GHz 24 GHz -55 dBc - -
13 GHz 26 GHz -55 dBc - -
18. Set the Spectrum Analyzer as follows:
PRESET
COUPLING DC
REF LVL +10 dBm
EXT REF On
FREQ 4.399994 GHz
SPAN 2 MHz