96000 Series
Service Manual
3-74
19. Set the UUT as follows:
Mode Leveled Sine
Signal Microwave Output
Frequency 4.399994 GHz
Level +4.6 dBm
Output OPER
20. Set the Spectrum Analyzer to single sweep, initiate a sweep and wait for the
sweep to complete. Push MARKER
PEAK.
21. Using the spectrum analyzer marker delta functions, measure the amplitude
of any spurious signals detected. Verify any spurious signals at offsets
>±3 kHz from the output frequency and within the range of ±1 MHz are within
the listed limits.
22. Set the output frequency of the UUT and the spectrum analyzer center
frequency to each of the frequencies listed in Table 3-30, repeating steps 20
and 21.
23. Set the UUT output to STBY.
Table 3-30. Sub Harmonics and Interharmonics Test, Microwave Output
Output Frequency Sub/Interharmonic Frequency Limit
4.05 GHz 6.075 GHz - - < -60 dBc
6.20 GHz 9.300 GHz - - < -60 dBc
8.30 GHz 4.150 GHz 6.225 GHz 12.450 GHz < -60 dBc
12.00 GHz 3.000 GHz 6.000 GHz 9.000 GHz < -60 dBc
16.00 GHz 4.000 GHz 8.000 GHz 12.000 GHz < -60 dBc
18.50 GHz 4.625 GHz - < -60 dBc
22.00 GHz 5.500 GHz - - < -60 dBc
27.00 GHz 6.750 GHz - - < -60 dBc
Note
The Microwave output signals at 4.024 GHz use the same
frequency generation and synthesis circuits that are tested during
the Leveling Head output spurious content tests. For signal outputs
at >4.024 GHz frequency multiplication techniques are used, and no
additional spurious generation mechanisms are present. Therefore,
Microwave output spurious content testing is not required to
adequately verify Product performance.