Performance Test Record
Performance Test Record A
A-7
Level Accuracy - 50
Ω
Levelling Head (96040A and 96270A)
Use an AC Measurement Standard, a power meter and power sensors, and a
high performance spectrum analyzer/measuring receiver to make the
measurements shown in Tables A-16 and A-17.
Table A-12. Level Accuracy (50 Ω) Test
Level Frequency Actual Tolerance
+16 dBm
1 kHz ±0.03 dB
20 kHz ±0.03 dB
100 kHz ±0.03 dB
10 MHz ±0.05 dB
100 MHz ±0.05 dB
300 MHz ±0.07 dB
1.4 GHz ±0.2 dB
+13 dBm
1 kHz ±0.03 dB
20 kHz ±0.03 dB
100 kHz ±0.03 dB
10 MHz ±0.05 dB
100 MHz ±0.05 dB
300 MHz ±0.07 dB
1.4 GHz ±0.2 dB
3 GHz ±0.3 dB
4 GHz ±0.3 dB
+3 dBm
1 kHz ±0.03 dB
20 kHz ±0.03 dB
100 kHz ±0.03 dB
10 MHz ±0.05 dB
100 MHz ±0.05 dB
300 MHz ±0.07 dB
1.4 GHz ±0.2 dB
3 GHz ±0.3 dB
4 GHz ±0.3 dB