A-6
Frequency Counter Accuracy (96040A and 96270A)
Use a frequency standard to make the measurements shown in Tables A-10 and
A-11.
Table A-10. Frequency Counter Accuracy Test
Frequency Actual
Tolerance
(After 24 hour warm-up)
10 MHz ±0.051 ppm
Table A-11. Alternative Frequency Counter Functional Test
Frequency Actual Tolerance
10 MHz ±1 digit