Performance Test Record
Performance Test Record A
A-5
Modulation (96040A and 96270A, Optional)
Use a high performance measuring receiver to make the measurements shown
in Tables A-6, A-7, A-8, and A-9.
Table A-6. Amplitude Modulation Rate Test
Level Frequency
Modulation
Rate
Depth Actual Tolerance
+10 dBm 30 MHz
20 kHz 50 % ±1 Hz
100 kHz 50 % ±10.0 Hz
Table A-7. Amplitude Modulation Depth Test
Level Frequency
Modulation
Rate
Depth Actual Tolerance
+14 dBm
125 MHz
1 kHz 90 % ±2.5 %
100 kHz 90 % ±2.5 %
1 GHz
1 kHz 90 % ±2.5 %
20 kHz 90 % ±2.5 %
Table A-8. Frequency Modulation Rate Test
Level Frequency
Modulation
Rate
Deviation Actual Tolerance
+13 dBm
62.5 MHz 400 Hz 30 kHz ±0.11 Hz
62.5 MHz 300 kHz 1 MHz ±10 Hz
Table A-9. Frequency Modulation Deviation Test
Level Frequency
Modulation
Rate
Deviation Actual Tolerance
+13 dBm
125 MHz
400 Hz 10 kHz ±300 Hz
100 kHz 10 kHz ±300 Hz
400 Hz 300 kHz ±9.0 kHz
100 kHz 300 kHz ±9.0 kHz
1 GHz
400 Hz 1 MHz ±30 kHz
300 kHz 1 MHz ±30 kHz