Table 12. Level Accuracy (50 Ω) Test (cont).
Level Frequency Actual Tolerance
-75 dBm
100 kHz ±0.5 dB
10 MHz ±0.1 dB
100 MHz ±0.1 dB
300 MHz ±0.3 dB
1.4 GHz ±0.5 dB
3 GHz ±1.0 dB
4 GHz ±1.0 dB
-85 dBm
100 kHz ±0.5 dB
10 MHz ±0.3 dB
100 MHz ±0.3 dB
300 MHz ±0.5 dB
1.4 GHz ±1.0 dB
3 GHz ±1.0 dB
4 GHz ±1.0 dB
-95 dBm
10 MHz ±0.7 dB
100 MHz ±0.7 dB
300 MHz ±1.5 dB
1.4 GHz ±1.5 dB
Table A-13. Optional Ultra-Low Level Accuracy (50 Ω) Test
Level Frequency Actual Tolerance
-105 dBm
10 MHz ±0.7 dB
100 MHz ±0.7 dB
300 MHz ±1.5 dB
1.4 GHz ±1.5 dB
-115 dBm
10 MHz ±0.7 dB
100 MHz ±0.7 dB
300 MHz ±1.5 dB
1.4 GHz ±1.5 dB
-124 dBm
10 MHz ±0.7 dB
100 MHz ±0.7 dB
300 MHz ±1.5 dB
1.4 GHz ±1.5 dB