EasyManua.ls Logo

Fluke 96000 Series - Page 181

Fluke 96000 Series
198 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Performance Test Record
Performance Test Record A
A-11
Attenuation Accuracy - 50
Ω
Levelling Head (96040A and 96270A)
Use a high performance measuring receiver to make the measurements shown
in Table A-14 relative to +10 dBm.
Table A-14. Attenuation Accuracy (50 Ω) Test
Frequency Attenuation Actual Specification Uncertainty Tolerance
10 MHz
0 dB --- Reference Reference Reference
-7 dB ±0.02 dB
-17 dB ±0.02 dB
-27 dB ±0.02 dB
-37 dB ±0.02 dB
-47 dB ±0.02 dB
-57 dB ±0.03 dB
-67 dB ±0.05 dB
-76 dB ±0.07 dB
-85 dB ±0.07 dB
-95 dB ±0.15 dB
-105 dB ±0.15 dB
50 MHz
0 dB --- Reference Reference Reference
-7 dB ±0.02 dB
-17 dB ±0.02 dB
-27 dB ±0.02 dB
-37 dB ±0.02 dB
-47 dB ±0.02 dB
-57 dB ±0.03 dB
-67 dB ±0.05 dB
-76 dB ±0.07 dB
-85 dB ±0.07 dB
-95 dB ±0.15 dB
-105 dB ±0.15 dB

Table of Contents