EasyManua.ls Logo

Fluke 96000 Series - Page 182

Fluke 96000 Series
198 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
96000 Series
Users Manual
A-12
Table A-14. Attenuation Accuracy (50 Ω) Test (cont.)
Frequency Attenuation Actual Specification Uncertainty Tolerance
100 MHz
0 dB --- Reference Reference Reference
-7 dB ±0.02 dB
-17 dB ±0.02 dB
-27 dB ±0.02 dB
-37 dB ±0.02 dB
-47 dB ±0.02 dB
-57 dB ±0.03 dB
-67 dB ±0.05 dB
-76 dB ±0.07 dB
-85 dB ±0.07 dB
-95 dB ±0.15 dB
-105 dB ±0.15 dB
VSWR - 50
Ω
Levelling Head (96040A and 96270A, Optional)
Use a directional bridge and a spectrum analyzer to make the measurements
shown in Table A-15.
Table A-15. VSWR (50 Ω) Test
Level Frequency Actual Tolerance
+13 dBm
100 MHz <1.05
2 GHz <1.10
3 GHz <1.15
4GHz <1.20
+3 dBm
100 MHz <1.05
2 GHz <1.10
3 GHz <1.15
4GHz <1.20
-7 dBm
100 MHz <1.05
2 GHz <1.10
3 GHz <1.15
4GHz <1.20

Table of Contents