A-14
Table A-16. Level Accuracy (75 Ω) Test (cont.)
Level Frequency Actual Tolerance
-13 dBm
1 kHz ±0.12 dB
20 kHz ±0.12 dB
100 kHz ±0.12 dB
10 MHz ±0.12 dB
100 MHz ±0.12 dB
300 MHz ±0.15 dB
1.4 GHz ±0.25 dB
2 GHz ±0.3 dB
-23 dBm
1 kHz ±0.12 dB
20 kHz ±0.12 dB
100 kHz ±0.12 dB
10 MHz ±0.12 dB
100 MHz ±0.12 dB
300 MHz ±0.15 dB
1.4 GHz ±0.25 dB
2 GHz ±0.3 dB
-33 dBm
1 kHz ±0.15 dB
20 kHz ±0.15 dB
100 kHz ±0.15 dB
10 MHz ±0.15 dB
100 MHz ±0.15 dB
300 MHz ±0.2 dB
1.4 GHz ±0.5 dB
2 GHz ±0.5 dB
-43 dBm
1 kHz ±0.15 dB
20 kHz ±0.15 dB
100 kHz ±0.15 dB
10 MHz ±0.15 dB
100 MHz ±0.15 dB
300 MHz ±0.2 dB
1.4 GHz ±0.5 dB
2 GHz ±0.5 dB