EasyManua.ls Logo

Fluke 96000 Series - Page 67

Fluke 96000 Series
198 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Performance Tests
Performance Test 3
3-25
Table 3-12. 50 Ω Leveling Head, High Frequency Level Accuracy Test Points, Diode Power Sensor
(96040A and 96720A)
Amplitude Frequency Reading P
out
Tolerance
-7 dBm
100 kHz P
2
Reference
10 MHz P
3
±0.05 dB
100 MHz P
3
±0.05 dB
300 MHz P
3
±0.07 dB
1.4 GHz P
3
±0.20 dB
3 GHz P
3
±0.30 dB
4 GHz P
3
±0.30 dB
-17 dBm
100 kHz P
2
Reference
10 MHz P
3
±0.05 dB
100 MHz P
3
±0.05 dB
300 MHz P
3
±0.07 dB
1.4 GHz P
3
±0.20 dB
3 GHz P
3
±0.30 dB
4 GHz P
3
±0.30 dB
-27 dBm
100 kHz P
2
Reference
10 MHz P
3
±0.05 dB
100 MHz P
3
±0.05 dB
300 MHz P
3
±0.07 dB
1.4 GHz P
3
±0.20 dB
3 GHz P
3
±0.30 dB
4 GHz P
3
±0.50 dB
-37 dBm
100 kHz P
2
Reference
10 MHz P
3
±0.05 dB
100 MHz P
3
±0.05 dB
300 MHz P
3
±0.07 dB
1.4 GHz P
3
±0.20 dB
3 GHz P
3
±0.30 dB
4 GHz P
3
±0.50 dB
-47 dBm
100 kHz P
2
Reference
10 MHz P
3
±0.05 dB
100 MHz P
3
±0.05 dB
300 MHz P
3
±0.07 dB
1.4 GHz P
3
±0.20 dB
3 GHz P
3
±0.30 dB
4 GHz P
3
±0.50 dB

Table of Contents