High Frequency Test Points (steps 8-17)
8. Connect the 75
Ω Leveling Head to the 75 Ω Power Sensor via a precision
75
Ω Type-N female-to-female adapter. See Figure 3-11 for equipment
connections.
Instrument
Power Meter
Power Sensor
75 Leveling
Head
N Adapter
hpn215.eps
Figure 3-11. Equipment Connections- 75 Ω Leveling Head, High Frequency Level Accuracy Test
Points (96040A and 96270A)
9. Set the UUT as follows:
Mode Leveled Sine
Frequency 100 kHz
Level +10 dBm
Output OPER
10. Configure the Power Meter to indicate readings in dBm. Enter the
measurement frequency (0.1 MHz) into the Power Meter to enable cal factor
(frequency) correction.