Performance Tests
Performance Test 3
3-39
Table 3-17. Level Accuracy Test (75 Ω), Low Frequency Test Points
Amplitude Frequency Tolerance
+10 dBm
1 kHz ±0.12 dB
20 kHz ±0.12 dB
100 kHz (P
1
) ±0.12 dB
+7 dBm
1 kHz ±0.12 dB
20 kHz ±0.12 dB
100 kHz (P
1
) ±0.12 dB
-3 dBm
1 kHz ±0.12 dB
20 kHz ±0.12 dB
100 kHz (P
1
) ±0.12 dB
-13 dBm
1 kHz ±0.12 dB
20 kHz ±0.12 dB
100 kHz ±0.12 dB
-23 dBm
1 kHz ±0.12 dB
20 kHz ±0.12 dB
100 kHz (P
1
) ±0.12 dB
-33 dBm
1 kHz ±0.15 dB
20 kHz ±0.15 dB
100 kHz (P
1
) ±0.15 dB