Performance Tests
Performance Test 3
3-45
Table 3-19. 75 Ω Leveling Head, Low Level Accuracy Test Points, Part A (96040A and 96270A)
Part Frequency Amplitude Reading P
out
Tolerance
A
100 kHz
-33 dBm Reference Reference
-53 dBm P
rel
±0.15 dB
-72 dBm P
rel
±0.20 dB
-81 dBm (P
-81
) P
rel
±0.70 dB
10 MHz
-33 dBm Reference Reference
-53 dBm P
rel
±0.15 dB
-72 dBm P
rel
±0.20 dB
-81 dBm (P
-81
) P
rel
±0.70 dB
100 MHz
-33 dBm Reference Reference
-53 dBm P
rel
±0.15 dB
-72 dBm P
rel
±0.20 dB
-81 dBm (P
-81
) P
rel
±0.70 dB
300 MHz
-33 dBm Reference Reference
-53 dBm P
rel
±0.20 dB
-72 dBm P
rel
±0.20 dB
-81 dBm (P
-81
) P
rel
±0.70 dB
1.4 GHz
-33 dBm Reference Reference
-53 dBm P
rel
±0.15 dB
-72 dBm P
rel
±0.20 dB
-81 dBm (P
-81
) P
rel
±1.00 dB
2 GHz
-33 dBm Reference Reference
-53 dBm P
rel
±0.50 dB
-72 dBm P
rel
±0.50 dB
-81 dBm (P
-81
) P
rel
±1.00 dB