96000 Series
Service Manual
3-46
Table 3-20. 75 Ω Leveling Head, Low Level Accuracy Test Points, Part B (96040A and 96270A)
Part Frequency Amplitude Reading P
out
Tolerance
B
100 kHz
-33 dBm Reference Reference
-53 dBm P
rel
±0.15 dB
-63 dBm P
rel
±0.20 dB
-81 dBm P
rel
±0.70 dB
10 MHz
-33 dBm Reference Reference
-53 dBm P
rel
±0.20 dB
-63 dBm P
rel
±0.20 dB
-81 dBm P
rel
±0.70 dB
-101 dBm (P
-101
) P
rel
±1.50 dB
125 MHz
-33 dBm Reference Reference
-53 dBm P
rel
±0.15 dB
-63 dBm P
rel
±0.20 dB
-81 dBm P
rel
±0.70 dB
-101 dBm (P
-101
) P
rel
±1.50 dB
300 MHz
-33 dBm Reference Reference
-53 dBm P
rel
±0.15 dB
-63 dBm P
rel
±0.20 dB
-81 dBm P
rel
±0.70 dB
-101 dBm (P
-101
) P
rel
±1.50 dB
1.4 GHz
-33 dBm Reference Reference
-53 dBm P
rel
±0.50 dB
-63 dBm P
rel
±0.50 dB
-81 dBm P
rel
±1.00 dB
-101 dBm (P
-101
) P
rel
±1.50 dB
2 GHz
-33 dBm Reference Reference
-53 dBm P
rel
±0.50 dB
-63 dBm P
rel
±0.50 dB
-81 dBm P
rel
±1.00 dB
-101 dBm (P
-101
) P
rel
±1.50 dB

Table of Contents