BT521
Users Manual
3-12
Measure AC Voltage
The Product supplies two independent readings to show the rms and frequency of ac
voltage.
To measure ac voltage, turn the rotary switch to . See Figure 3-4 for connections.
Vac
UPS
hsz021.eps
Figure 3-4. Measure AC Voltage
Measurement Range
The ac voltage measurement has only one range: 600 V. The frequency measurement
uses auto ranging. This cannot be changed.
Save AC Voltage Readings
In Meter mode, push the Save softkey to save the current ac voltage reading and the
timestamp. All saved data is stored in chorological order.
In Sequence mode, push the Save softkey to save the current ac voltage reading and
the timestamp. The current battery number increases 1 automatically. The cell that
corresponds to the tested battery becomes full, and the cursors move forward.
Note
Up to 20 ac voltage readings can be saved in one profile.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Fluke BT521 and is the answer not in the manual?

Fluke BT521 Specifications

General IconGeneral
CertificationRoHS, IEC 61326
Product colorBlack, Yellow
Frequency range50 - 60 Hz
Maximum measured voltage600 V
Minimum measured voltage6 V
International Protection (IP) codeIP40
Operating temperature (T-T)0 - 40 °C
Display typeLCD
Built-in displayYes
Battery voltage7.4 V
Weight and Dimensions IconWeight and Dimensions
Depth58 mm
Width103 mm
Height220 mm
Weight850 g

Related product manuals