EasyManua.ls Logo

Fluke PM6685 - Page 40

Fluke PM6685
94 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
n
Prescaler 3.0 GHz, PM9624
See Chapter 2, Performance Check, for verification.
This prescaler cannot be repaired in a local workshop. It must be sent
to your Fluke representative for repair.
GPIB Interface and Analog Output
n
General Remark
If the GPIB board is suspected to be faulty, be sure the basic instru
-
ment is OK by performing a few functional checks after the ribbon
cable has been disconnected from J18.
n
Analog Output
The microcontroller generates a PWM signal that is applied to pin 1
on U101. The frequency is approximately 20 Hz, but the duty cycle is
dependent on several factors like the frequency of the measured sig
-
nal, the measurement time, and the selected scaling factor.
The PWM signal is converted to a DC voltage between 0 V and
4.98 V by integration, first in a passive RC network (R101, C103,
R102, C102 and then in an active integrator U103.
Setup
Connect the counter to line power.
Switch on the counter.
Press PRESET and then ENTER.
Connect a DMM to the BNC output BU102.
Activate the analog output.
Select AUX MENU.
Press DATA ENTRY p/q until the display reads ANA
-
LOG OUT.
Press ENTER
Press DATA ENTRY p/q to select ON.
Press ENTER.
Press DATA ENTRY p/q until the display reads 1.0
-3
V.
Press ENTER.
Connect a LF synthesizer to Input A on the counter.
Set the synthesizer to 500 Hz, 1 V
PP
Read the DMM result. The voltage should be 2.49 V ± 35 mV.
Minor deviations can depend on the settings of the trimmer potenti
-
ometers for ZERO and FULL SCALE. See Chapter 6, Calibration
Adjustments, for a decription of the procedure to follow.
Large deviations indicate a fault. Trace the signal through the inte
-
gration chain with traditional troubleshooting techniques and replace
defective circuits. The duty cycle at U101:1 should be 50 %. Also re
-
fer to GPIB Interface Including Analog Output in Chapter 4, Circuit
Descriptions.
n
Bus Interface
A simple method to check the most fundamental functions of the in
-
terface is to send the standardized query message
*
IDN? and check
the response string.
Setup
Make sure you have access to a PC with GPIB capability.
Check that there is a program installed that can send simple
commands entered via the keyboard and that can receive and
display the response strings.
Connect the GPIB connectors of the counter and the PC by
means of a standard GPIB cable.
Set the address switches on the counter (the five rightmost
ones seen from the rear) so that their binary weight corre
-
sponds to the wanted decimal value between 0 and 30.
Send the command
*
IDN? to the counter and observe the re
-
sponse string. See the programming manual for more
information on the response format and contents.
You can also try the command
*
OPT? to get a listing of in
-
stalled options (except OCXO).
If you find a fault, continue with traditional troubleshooting tech
-
niques and replace defective circuits. Try to exercise the address/data
bus by writing small program loops. Look for stuck nodes with an
oscilloscope.
5-8 Troubleshooting
0
Sensitivity
-10 dBm
-20 dBm
-30 dBm
-40 dBm
-50 dBm
1GHz 2GHz
2.5 GHz
Frequency
Fig. 5-6 Specified and typical sensitivity of input C
(PM9624).
U109
U101
U106
BU101
U103
Zero
GPIB
BU102
BU103
Full Scale
U107
U111
U108
U113
U116
U117
U114 U115
Fig. 5-7 Component layout, GPIB interface.

Other manuals for Fluke PM6685