EasyManua.ls Logo

Flybox Vigilus - Page 55

Default Icon
92 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
VIGILUS - Operating manual
Flybox
55
Rev. 1.0
®
Fuel level sensors calibration
Left Tank / Main Tank / Right Tank: Choose which tank you
want to calibrate. For example if you select the left tank this
screen will appears:
If you’ve never performed the
calibration before for the selected
tank, you will see the “Calibration
never done!” indication.
Press the “START” button to start the calibration; the display
will shows this screen:
(2) Calibration step
to be performed
(3) Electrical output
of the fuel level
sensor
(1) Action to be
performed
Step#01 (EMPTY TANK): Drain the tank such that only the
unusable fuel remain in the tank. Wait until the indication (3)
is stable and then click the knob [NEXT] to confirm and go to
the next step.
Step#02: Add to the tank the indicated fuel quantity (it's the
same quantity choosed with the “Cal. steps” parameter), on this
example it's required to add 2 liters of fuel:
Instrument configuration