1. S
witch main switch to "0".
2. S
witch the device on at the main switch while holding down the
Stop key.
The device is now in "Setup" mode.
3. The "Sie
ving Time" display now shows the default seng "04" for
230 V, "06" for 115 V. The value "07" is preset for a mains voltage
of 100 V (in Japan).
4. These v
alues can be changed with the "Sieving Time" +/- key and
adapted to the mains voltage.
n 230 V - 04
n 115 V - 06
n 100 V - 07
Save the set values by clicking the stop key.
5. Aer making adjustments, switch on the main switch again.
4.5.1.2
Version PRO (valid from serial number 06000) and Version SPARTAN (valid from serial number 05000)
CAUTION!
If the v
alue 01 is set at a voltage of 230V~, it will result
in a defect in the sieve shaker during operaon. Failure
to observe this will render void the guarantee, and
releases us from liability for any resulng damage to
the device.
NOTICE!
Only qualied personnel may change the voltage range
on the device!
1. S
witch main switch to "0".
2. S
witch the device on at the main switch while holding down the
Stop key.
The device is now in "Setup" mode.
3. The "Sie
ving Time" display now shows the default seng "02" for
230 V, "01" for 115 V.
4. These v
alues can be changed with the "Sieving Time" +/- key and
adapted to the mains voltage.
n 230 V - 02
n 115 V - 01
Save the set values by clicking the stop key.
5. Aer making adjustments, switch on the main switch again.
Installaon