6.1.3 Feed quanty for dry or wet sieving
Maximum load:
‐ Sie
ving stock < 1 kg
‐ Sieves and sieving stock < 6 kg
6.2 Dry sieving
For dry sieving, use the dry sieve pan (without outlet) and the dry
clamping lid (plexiglas clamping lid without nozzles).
6.2.1 Sieving parameters
Parameters Coarse sample Fine sample
Sieving me 3...20 min 15....30 min
Amplitude 2.5....3 mm 1.5....2.5 mm
In order to achieve an amplitude set point of 3 mm (full deecon), at
leas
t 3 sieves, collecng vessels and clamping lids must be clamped on
the sieve shaker. If fewer sieves are clamped, it may not be possible to
achieve an amplitude set point of 3 mm.
(Sieves with 200 mm diameter)
Using the device
- 25 -