6.4 Micro-precision sieving with the ANALYSETTE 3 PRO
6.4.1
Sieving parameters
Parameters 1 micro sieve Max. 4 micro sieves
Sieving me 20...30 min 60 min
Amplitude in "MICRO"
operang mode
0.1...0.2 mm 0.5 mm
6.4.2 Feed quanty
Depending on parcle size and sieve aperture size, it is possible to sieve a
ma
ximum of 200 mg (up to 1 g for coarse sieves of 50 - 100 µm). Deter-
mine the suitable feed quanty through experiment.
NOTICE!
The de
vice must be in "MICRO" operang mode. The
"MICRO" key must light up, otherwise there is a risk of
destroying the sieves.
In the "MICRO" operang mode the amplitude is limited, so that only
values between 0.1 and 0.5 mm can be selected.
NOTICE!
For sieving with micro-precision sieves, the toothed
belts for dry or wet sieving should be removed from the
belt clamping device. (See
Ä
Chapter 6.1.1 „Fing and
clamping the sieves“ on page 23).
6.4.3 Fing the micro-precision sieves
NOTICE!
Clamp a ma
ximum of 4 micro-precision sieves!
Using the device
- 29 -