6.4 Micro-precision sieving with the ANALYSETTE 3 PRO
6.4.1
Sieving parameters
Parameters 1 micro sieve Max. 4 micro sieves
Sieving me 20...30 min 60 min
Amplitude in "MICRO"
operang mode
0.1...0.2 mm 0.5 mm
6.4.2 Feed quanty
Depending on parcle size and sieve aperture size, it is possible to sieve a
ma
ximum of 200 mg (up to 1 g for coarse sieves of 50 - 100 µm). Deter-
mine the suitable feed quanty through experiment.
NOTICE!
The de
vice must be in "MICRO" operang mode. The
"MICRO" key must light up, otherwise there is a risk of
destroying the sieves.
In the "MICRO" operang mode the amplitude is limited, so that only
values between 0.1 and 0.5 mm can be selected.
NOTICE!
For sieving with micro-precision sieves, the toothed
belts for dry or wet sieving should be removed from the
belt clamping device. (See
Ä
Chapter 6.1.1 „Fing and
clamping the sieves“ on page 23).
6.4.3 Fing the micro-precision sieves
NOTICE!
Clamp a ma
ximum of 4 micro-precision sieves!
Using the device
- 29 -