EasyManua.ls Logo

FUTABA Mold Marshalling System MPS08 - Sensor Calibration and Quick Check

FUTABA Mold Marshalling System MPS08
46 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
37
7-3 Checking the Operation of Sensor
7-3-1 Sensor calibration
If waveforms are not displayed or the pressure value is too low (or too high) and you think that a pressure
sensor is faulty, it is recommended to calibrate the sensor. Sensor calibration is available for a fee.
Please contact the nearest FUTABA sales office (written on the back cover).
7-3-2 Quick sensor checking
Follow the procedure below to simply check the operation of a sensor.
Press MAINTENANCE MODE. Measurement starts
Press the ZERO RESET button.
Press the tip of the sensor to check whether waveforms are displayed.
(This is only a simple test to check whether the sensor works.)
* The output of the sensor is forcibly reset to 0.
When pressing “ZERO RESET”, make sure that no
force (load) is applied to the sensor.
* When applying a load to the tip of the sensor, use a soft
object, etc. so as not to damage the sensor or mold
(scratches, corrosion etc.).
* When applying a load to the tip of the sensor, never
apply a load exceeding the ratings or a shock load. It
could damage the sensor.
* Perform a simple operation check at the customers
own risk.

Table of Contents