GE Healthcare
Life Sciences
ÄKTAprocess™
Operating Instructions
Original instructions

Table of Contents

Other manuals for GE AKTAprocess

Questions and Answers

  • G
    Gina NguyenSep 7, 2025
    What to do if no system is found when starting up UNICORN on GE AKTAprocess?
    • J
      John MossSep 7, 2025
      If no system is found when starting up UNICORN, ensure the system is switched on. Check the communication cable and connectors. Reboot the PC: shut down Windows, switch off the SYSTEM POWER SWITCH, wait at least 5 seconds, and then restart the system.
  • N
    Nicole HigginsAug 22, 2025
    What to do if pump is not working in GE AKTAprocess?
    • K
      Kayla VegaAug 22, 2025
      If the pump is not working, check if the emergency button has been pressed. Also, verify the air supply pressure and adjust it if needed. Inspect the method and valves. Try entering Pump:ManFlow > 1%. If the issue persists, contact GE Healthcare service personnel.
  • M
    Monica SpencerSep 5, 2025
    How to troubleshoot blockage or partial blockage of flow path in GE AKTAprocess?
    • A
      Ann MezaSep 5, 2025
      If there is a blockage or partial blockage of the flow path, flush through to clear the blockage.
  • J
    Juan PowellSep 3, 2025
    What to do if the actual flow rate deviates from the displayed flow rate on GE AKTAprocess Laboratory Equipment?
    • C
      Courtney MaysSep 3, 2025
      If the actual flow rate deviates from the displayed flow rate, or a flow rate is displayed when the pumps are off, this can be caused by changes in running conditions. Perform a zero-point adjustment of the flow meter to resolve this.
  • B
    Brian RossSep 2, 2025
    How to fix noisy UV-signal, signal drift or instability in GE AKTAprocess?
    • H
      Heather MontgomerySep 2, 2025
      Noisy UV-signal, signal drift, or instability can be caused by bad UV fiber connections. Check the connections of the UV cell optical fiber and replace them if necessary.
  • J
    Jennifer BrownAug 31, 2025
    How to fix compressed air failure during a run in GE Laboratory Equipment?
    • D
      deannahowellAug 31, 2025
      If there is a compressed air failure during a run, it may be caused by insufficient air pressure due to malfunctioning air supply equipment or a leaking air hose. Rectify the air supply problem and restart the run by pressing the Continue button.
  • S
    Stephen JohnsonAug 30, 2025
    What to do if there is too high outlet pressure in GE AKTAprocess Laboratory Equipment?
    • N
      Noah OrtegaAug 30, 2025
      If there's too high outlet pressure, verify that the connected outlet is actually in use and that outlet containers are not placed too high relative to the pump. Inspect the tubing to the outlet container for excessive length, small internal diameter, narrow sections, or blockages. Also, check for any non-functioning valves or a malfunctioning valve at the container battery limit.
  • K
    Kayla SmithAug 29, 2025
    What to do if there is little or no flow in GE Laboratory Equipment?
    • O
      olsendustinAug 29, 2025
      If you experience little or no flow, check that the connected inlet is actually in use. Ensure inlet containers are not placed too low relative to the pump. Verify that there is liquid supplied to the pump. Also, check the tubing from the inlet container for issues like excessive length, small internal diameter, narrow sections, or blockages. A malfunctioning valve at the container battery limit could also be the cause.
  • R
    riverabrandonAug 27, 2025
    What to do if there is erratic flow, noisy baseline signal, or irregular pressure trace in GE Laboratory Equipment?
    • M
      Mark CastilloAug 27, 2025
      Erratic flow, a noisy baseline signal, or an irregular pressure trace can be caused by gas bubbles passing through or trapped in the pump. Ensure there is a sufficient supply of liquid and check all connections for leaks.
  • K
    Kara ThomasAug 26, 2025
    What to do if the alarm 'Valve error' is displayed on GE AKTAprocess?
    • B
      barkertoddAug 27, 2025
      If the 'Valve error' alarm is displayed, perform a valve calibration. If the calibration doesn't restore the valve function, contact your local GE Healthcare representative. Note that if the 'Valve error' alarm is accompanied by 'No air supply to the system', the alarm is caused by a compressed air failure.

Summary

Introduction

About this manual

Defines the manual's purpose, scope, and typographical conventions.

Important user information

Crucial information before operating, intended use, and prerequisites for safe use.

Regulatory information

Details manufacturing, CE, international, additional, and environmental compliance.

Safety instructions

Safety precautions

Covers general, personal, installation, power, and operational safety measures.

Labels

Explains the meaning of various system and safety labels found on the equipment.

Emergency procedures

Details steps for emergency shutdown, power failure response, and system restart.

System description

Configurations

Summarizes standard/optional components, tubing, and cabinet types.

Illustrations of ÄKTAprocess

Provides front and rear views of standard and optional system configurations.

Standard components

Details on structural, control, inlets/outlets, meters, pump, air trap, and valves.

UNICORN control system

Describes software modules, workflow, manual instructions, indicators, and alarms.

Installation

Site requirements

Details requirements for space, floor load, ambient environment, power, and air supply.

Transport

Provides guidelines for transporting the system in crate and when unpacked.

ÄKTAprocess Setup

Covers assembly, control system setup, connections, and column connection procedures.

Power supply

Details power requirements, UPS, grounding, circuit breakers, and cable specifications.

Operation

Prepare the system

Steps for starting system/software, preparing components, testing, and final checks.

Perform a run

Instructions for starting, monitoring, and ending a system run.

Shut down the system and software

Procedures for safely shutting down the system and the UNICORN software.

Maintenance

User maintenance schedule

Lists weekly, monthly, and annual maintenance tasks for system components.

Cleaning

Procedures and recommendations for cleaning, including CIP and sanitization.

Storage

Provides procedures for short-term and long-term storage of the system.

Calibration of pH probe

Details required equipment and steps for calibrating the pH probe.

Troubleshooting

System

Addresses common system problems like PC connection and power failure.

Pump

Troubleshooting for pump not working, low flow, or high outlet pressure.

UV curve

Symptoms and corrective actions for issues with UV signals or drift.

Reference information

Specifications

Contains general technical data, cabinet/system specs, and temperature/pressure limits.

Chemical resistance

Lists allowed exposure times and concentrations for various chemicals.

UNICORN method for system CIP;Sanitization;Rinsing

Provides an example UNICORN method for system cleaning procedures.

GE AKTAprocess Specifications

General IconGeneral
BrandGE
ModelAKTAprocess
CategoryLaboratory Equipment
LanguageEnglish