September 2004
2-30 Programming Site Data
Accessing a Totalizer/
Frequency Output
These outputs can be configured for frequency or totalizer output
using PanaView™.
The totalizer output issues one pulse per selected volume of flow. The
meter produces a pulse each time the programmed amount of flow
passes through the pipe.
The frequency output issues a frequency that is proportional to the
assigned measurement parameter, such as flow rate.
Output B and D can be configured independently. Typically, Output B
is used for Channel 1 and Output D is used for Channel 2; however,
outputs can be configured to measure either channel.
1. Press [ESC], [ENTER], [ESC]. The meter enters the user program.
2. Press [
X] until PROG appears and press [ENTER].
3. Press [
X] until GLOBL appears and press [ENTER].
4. Press [
X] until I/O appears and press [ENTER].
5. Press [
X] until OPTN appears and press [ENTER].
6. Press [
X] to select B or D and press [ENTER].
7. If you are configuring a
• frequency output, proceed to the next section Setting Up a
Frequency Output on page 2-31.
• totalizer output, proceed to Setting Up a Totalizer Output on
page 2-32.