7-6 170 Series Monitor Revision C
2000947-004
Serviceable Assemblies: Transducer Plug Replacement Kits
Figure 7-1. Tocotransducer Cable Preparation
1²
²
1
8
¤
1
²
1
2
¤
1²
yellow
#24 AWG Green Jumper Wire
#24 AWG Green Wire
²
1
8
¤
Oetiker Clamp
(Nautilus only)
Vent Tube (Nautilus only)
Shrink Tubing
Shield
Table 7-3. Tocotransducer Cable Wiring
Pin Number Pin Length Description Wire Color
1 short +PRESS Red
2 short –PRESS White
3 short
4 short +4 V Blue
5 short
6 short Reference/Enable #24 AWG Green Jumper and Yellow
7 LONG Shield Ground #24 AWG Green
8 short
9 short
10 short
11 short
12 short TOCO Enable #24 AWG Green Jumper