EasyManuals Logo

GE Corometrics 170 Series Service Manual

GE Corometrics 170 Series
248 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #137 background imageLoading...
Page #137 background image
Revision C 170 Series Monitor 7-5
2000947-004
Serviceable Assemblies: Transducer Plug Replacement Kits
Tocotransducer Cable Preparation
Refer to Figure 7-1 and Table 7-3 while following these steps:
1. Cut #24 AWG green wire (not supplied) to 1-
1
2 inches long and strip each end
1
8 inch. This will be used as a jumper wire.
2. Cut #24 AWG green wire to 2 inches long. Strip one end
3
4 inch.
3. Route the transducer cable through the gripper. Slide the gripper, followed by
the retaining ring, down the cable until they are out of the way.
4. For Nautilus Tocotransducers only: Slide Oetiker clamp over end of cable. This
clamp is not necessary for Trimline Tocotransducers.
5. Score around grey cable jacket 1 inch. Peel and remove jacket. Trim filler flush
to jacket.
6. For Nautilus Tocotransducers only: Cut vent tubing flush with jacket being
careful not to restrict air flow. The vent tubing is not present in Trimline cables.
7. Pull back the braided shield and twist into pigtail.
8. Solder the 2-inch prepared green wire to the shield, wrapping the
3
4-inch strip on to the shield. Cut wire flush with other conductors of the cable.
9. Install shrink tubing over the solder joint of the shield.
10. Strip all conductors
1
8 inch. Twist and tin conductors.
11. Locate the grounding contact pin (longer pin) in the kit. Solder this pin to the
shield wire.
12. Solder a short pin to the yellow wire and one end of the green jumper wire.
Both wires share this pin.
13. Solder a short pin to the free end of the green jumper wire.
14. Solder a short pin to each of the other conductors.
15. For Nautilus Tocotransducers only: Crimp Oetiker clamp over cable jacket
1
8
inch from end.
16. Continue with “Plug Assembly” on page 7-13.

Table of Contents

Other manuals for GE Corometrics 170 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE Corometrics 170 Series and is the answer not in the manual?

GE Corometrics 170 Series Specifications

General IconGeneral
BrandGE
ModelCorometrics 170 Series
CategoryMedical Equipment
LanguageEnglish

Related product manuals