8-4 Dashâ„¢ 3000/4000/5000 2000966-386D
Monitoring ECG
IntelliRate
WARNING
PATIENT RISK HAZARD—The heart rate reported by the
IntelliRate feature may reflect either the electrical or pulsatile heart
rate. An electrical heart rate may be reported on some pulseless
rhythms. A pulsatile heart rate may be reported by the IntelliRate
feature during ASYSTOLE if the patient is on a cardiac assist
device. In these situations, the monitor will not detect or call an
ASYSTOLE patient alarm. The user may elect to turn the IntelliRate
feature off for patients at risk of these events, otherwise patient
treatment may be delayed. Such patients should always be kept
under close observation.
The IntelliRate algorithm utilizes information contained in multiple physiologic
signals (for example, ECG, arterial blood pressure and pulse oximetry) to improve the
accuracy of the heart rate calculation.