Monitor Setup
2000966-386D Dash™ 3000/4000/5000 3-3
CARDIAC — Focuses on cardiac conductivity. Includes full arrhythmia analysis
and storage, adjustable ST measurement point and ST segment trending, storage
and templates.
CARDIO-PULMONARY — Focuses on cardiac and pulmonary hemodynamics.
Includes cardiac and pulmonary calculations, PA insert and wedge algorithms,
intra-aortic balloon pump algorithm and a thermodilution cardiac output
algorithm that includes predefined computation constants for major catheter
manufacturers.
To determine the software running on a monitor, select MORE MENUS >
MONITOR SETUP > SOFTWARE CONFIGURATION. The software package is
listed under ENABLED SOFTWARE OPTIONS.
For more information on the features and functionality of each software package,
refer to Appendix B.
Software Options
There are three software options. These options can be purchased separately and used
with any software package or in combination with other software options.
High Resolution Trends — Provides storage of up to 100 CRG events and up to
24 hours of CRG trend data.
12SL ECG — Provides an analysis program that includes the Gender Specific
Criteria and the Acute Cardiac Ischemia-Time Insensitive Predictive Instrument
(ACI-TIPI). It uses recorded ECG data to produce a numerical score which is the
predicted probability of acute cardiac ischemia. In addition, the gender specific
criteria improves the detection of acute myocardial infarctions (AMI) for adult
women under the age of 60.
Unity Network — Enables you to view other patients on the network, interface
with a central station and other network devices and perform COMBO or
ROVER COMBO monitoring.
To determine the software options running on a monitor, select MORE MENUS >
MONITOR SETUP > SOFTWARE CONFIGURATION. The software options are
listed under ENABLED SOFTWARE OPTIONS.
Monitoring/Admit Modes
There are four monitoring or admit modes. The mode determines how patients are
admitted to the monitor and how the monitor functions within a network
environment. It is defined during installation via the password protected SERVICE
MODE menu.
STANDARD — Permanently mount a monitor in a room. This mode does not
accommodate telemetry and does not require a network connection unless central
station communication is needed.
ROVER — Move or “rove” the monitor to the patient, rather than moving the
patient to a monitored room. This mode does not accommodate telemetry and
does not require a network connection unless central station communication is
needed.

Table of Contents

Other manuals for GE Dash 5000

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE Dash 5000 and is the answer not in the manual?

GE Dash 5000 Specifications

General IconGeneral
BrandGE
ModelDash 5000
CategoryMedical Equipment
LanguageEnglish

Related product manuals