EasyManuals Logo
Home>GE>Medical Equipment>Dash 5000

GE Dash 5000 Operator's Manual

GE Dash 5000
270 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #72 background imageLoading...
Page #72 background image
5-4 Dashâ„¢ 3000/4000/5000 2000966-386D
Managing Patients
Automatic Admit
The AUTO ADMIT feature defaults to ON, but can be turned off in MONITOR
DEFAULTS. When turned ON in MONITOR DEFAULTS, the AUTO ADMIT
feature provides automatic admission of the patient and activation of alarms. The
monitor will automatically admit the patient after 30 seconds of continuous ECG or
SpO
2
physiologic data has been detected. The capability to manually admit a patient
is still available at any time.
WARNING
The AUTO ADMIT feature is only available on Dash monitors
using software version 6.6 and later. For monitors earlier than
version 6.6, or if the ADMIT PATIENT TO ACTIVATE ALARMS
message continues to display on the monitor 30 seconds after ECG
or SpO
2
physiologic data has been detected, you must manually
admit the patient using the Admit/Discharge button on the keypad,
or through the ADMIT menu.
WARNING
A clinical site might consist of a mixed environment of Dash
monitors with auto admit turned ON and some with Auto Admit
turned OFF, or Dash monitors earlier than version 6.6 that do not
support the AUTO ADMIT feature. If the ADMIT PATIENT TO
ACTIVATE ALARMS message continues to display on the monitor
30 seconds after ECG or SpO
2
physiologic data has been detected,
you must manually admit the patient using the Admit/Discharge
button on the keypad, or through the ADMIT menu.
NOTE
Physiological ECG data is any condition that does not result in a LEADS FAIL
condition.
NOTE
Physiological SpO
2
data is any condition that does not result in a SPO2 PROBE
OFF or SPO2 PROBE DISCONNECT condition.
NOTE
The AUTO ADMIT feature is available in the OPERATING ROOM patient-
monitor type only when the monitor uses the COMBO monitoring mode.
To turn ON or turn OFF the Auto Admit feature, follow this procedure:
1. Select MORE MENUS > MONITOR SETUP > MONITOR DEFAULTS >
SETUP DEFAULT DISPLAY > AUTO ADMIT.
2. Select ON to enable the Auto Admit feature, or OFF to disable the Auto Admit
feature.

Table of Contents

Other manuals for GE Dash 5000

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE Dash 5000 and is the answer not in the manual?

GE Dash 5000 Specifications

General IconGeneral
BrandGE
ModelDash 5000
CategoryMedical Equipment
LanguageEnglish

Related product manuals