3-2 Dashâ„¢ 3000/4000/5000 2000966-386D
Monitor Setup
Getting Started
The main display has one menu option, MORE MENUS. The following menu
options are starting points to access and adjust the monitor’s settings.
Identifying Your Configuration
Software Packages
There are three software packages.
 BASIC — Includes standard-of-care parameters, lethal arrhythmia detection,
dose calculations and other features required by clinicians caring for acutely ill
patients.
Option Function
ALARM CONTROL View and modify alarm limits, change alarm levels and
adjust alarm volume. For more information, refer to Control
Settings on page 4-5.
VIEW OTHER PATIENTS View other GE monitors on the network. For more
information, refer to Chapter 5.
PATIENT DATA View patient data and access the calculation programs. For
more information, refer to Chapter 6.
MONITOR SETUP Access menu options to customize the monitor to meet your
needs. For more information, refer to Customizing Monitor
Setup on page 3-6.
ADMIT MENU Access the menu options to admit and discharge patients.
For more information, refer to Admit on page 5-2.
NOTE
The menu name changes in OPERATING ROOM
patient-monitor type to NEW CASE SETUP.
GO TO STANDBY
CAUTION
AUTO NBP MODE SHUT OFF—When you select the GO
TO STANDBY option or press the Standby key, the NBP
AUTO mode automatically shuts off.
NOTE
This option is only available in software version 6 or
later.
Select to temporarily suspend monitoring. All patient
monitoring, alarms and trending stop, until the Trim Knob
control, the Standby key or Admit/Discharge key is
pressed or the monitor is rebooted.
BATTERY STATUS Review battery information. For more information, refer to
Batteries on page A-7.