C-4 Dash™ 3000/4000/5000 2000966-386D
Factory Defaults
Parameter Limits
The following are the parameter alarm limit factory defaults. To access these options,
select MORE MENUS > MONITOR SETUP > MONITOR DEFAULTS > SETUP
DEFAULT LIMITS.
ADULT-ICU NEONATAL-ICU OPERATING ROOM
LOW HIGH LOW HIGH LOW HIGH
HR 50 150 90 200 –1 150
PVC/MIN 6 6 6
ST-I
–2.0 2.0 –2.0 2.0 –2.0 2.0
ST-II
ST-III
ST-V1
ST-AVL
ST-AVF
ST-AVR
ST-V2
ST-V3
ST-V4
ST-V5
ST-V6
NBP-S 80 200 40 100 40 200
NBP-D 20 120 20 60 20 120
NBP-M 40 140 30 70 40 140
ART-S 80 200 40 100 40 200
ART-D 20 120 20 60 20 120
ART-M 40 140 30 70 40 140
ART-R 50 150 90 200 –1 150
FEM-S 80 200 40 100 40 200
FEM-D 20 120 20 60 20 120
FEM-M 40 140 30 70 40 140
FEM-R 50 150 90 200 –1 150
UAC-S 80 200 40 100 40 200
UAC-D 20 120 20 60 20 120
UAC-M 40 140 30 70 40 140