EasyManuals Logo

GE Dash 5000 Operator's Manual

GE Dash 5000
270 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #25 background imageLoading...
Page #25 background image
Introduction
2000966-386D Dash™ 3000/4000/5000 1-11
CAUTION
MPSO —The use of a multiple portable socket outlet (MPSO) for a
system will result in an enclosure leakage current equal to the sum of
all individual earth leakage currents of the system if there is an
interruption of the MPSO protective earth conductor. Do not use an
additional extension cable with the MPSO as it will increase the
chance of the single protective earth conductor interruption.
CAUTION
NEGLIGENCE—GE does not assume responsibility for damage to
the equipment caused by improperly vented cabinets, improper or
faulty power, or insufficient wall strength to support equipment
mounted on such walls.
CAUTION
OPERATOR—Medical technical equipment such as this monitor/
monitoring system must only be used by persons who have received
adequate training in the use of such equipment and who are capable
of applying it properly.
CAUTION
POWER REQUIREMENTS— Before connecting the device to the
power line, check that the voltage and frequency ratings of the
power line are the same as those indicated on the unit's label. If this
is not the case, do not connect the system to the power line until you
adjust the unit to match the power source.
In U.S.A., if the installation of this equipment will use 240V rather
than 120V, the source must be a center-tapped, 240V, single-phase
circuit.
This equipment is suitable for connection to public mains as defined
in CISPR 11.
CAUTION
RESTRICTED SALE —U.S. federal law restricts this device to sale
by or on order of a physician.
CAUTION
SINGLE PATIENT USE —This equipment is designed for use on
one patient at a time. Using this equipment to monitor different
parameters on different patients at the same time compromises the
accuracy of data acquired.

Table of Contents

Other manuals for GE Dash 5000

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE Dash 5000 and is the answer not in the manual?

GE Dash 5000 Specifications

General IconGeneral
BrandGE
ModelDash 5000
CategoryMedical Equipment
LanguageEnglish

Related product manuals