5-12 Dash™ 3000/4000/5000 2000966-386D
Managing Patients
The host monitor is completely functional and two seconds of waveforms data
continue to display. The menus still control the host monitor and any changes you
make do not affect the function of the viewed monitor. Information windows will also
display over the viewed monitor.
Keep in mind the following guidelines when viewing a split view:
Admit a patient to view. If the monitor is discharged, the parameter windows
display (no waveforms) with a “DISCHARGED” message.
To maintain CO
2
monitoring on the host monitor while viewing another monitor,
change the position of CO
2
to the sixth position. The slower CO
2
sweep speed
will block all other waveforms adjacent or below the CO
2
position.
CO
2
waveform and parameter windows are not sent to the host monitor.
ST trends and complexes are not sent to the host monitor.
CRG+ display is not sent to the host monitor.
Removing a Viewed Monitor Display
To remove an alarming monitor display from the host monitor:
VOA — Select MAIN MENU or RETURN.
AVOA — Perform any one of the following actions:
Select MAIN MENU or RETURN.
Silence the alarming monitor at the host monitor or central station.
Resolve the alarm condition at the alarming monitor (all alarm levels except
CRISIS).
Silence the remote alarm. This also prevents new alarms from sounding,
even if a new alarm is of higher priority.
AutoView XM — Perform any one of the following actions:
Select MAIN MENU or RETURN.
Silence the alarming monitor at the host monitor or central station.
Resolve the alarm condition at the alarming monitor (all alarm levels except
CRISIS).
Silence the remote alarm when the host monitor’s AVOA SILENCE setting
is set to SILENCED. This also prevents new alarms from sounding, even if
a new alarm is of higher priority.
A more recent, higher priority alarm is received at the host monitor when the
host monitors AVOA SILENCE setting is set to NEW ALARM.
NOTE
To remove a non-alarming monitor display from the host monitor, select MORE
MENUS > VIEW OTHER PATIENTS > SELECT A BED TO VIEW. The
viewed bed is indicated with the word “VIEW”. Selecting that bed again will
remove from the host monitor.

Table of Contents

Other manuals for GE Dash 5000

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE Dash 5000 and is the answer not in the manual?

GE Dash 5000 Specifications

General IconGeneral
BrandGE
ModelDash 5000
CategoryMedical Equipment
LanguageEnglish

Related product manuals