IRECTION 5535208-100, REV. 2 LOGIQ E9 SERVICE MANUAL
3 - 74 Section 3-9 - Connectivity Setup and Tips
3-9-4-10 Spooler (cont’d)
Common Error Messages
No connection established – check network and log file: This usually indicates a bad
or missing network connection OR incorrect IP Address, AE Title, or Port # between the and service.
The remote system was not able to interpret the image information – check log
file: This indicates an issue with the image or patient data, such as a missing Patient ID.
The remote service is out of resources: For example, the PACS has no more storage space.
3-9-4-11 Wireless
Configuring a Wireless network is beyond the scope of this guide. Both the User and Service manuals
cover Wireless setup in detail. In particular, for setting up the WLAN, refer to the appropriate version the
Basic User Manual, Chapter 16, Configuring the Wireless Network (Option). (Available via the “Help” F1
key on the keyboard.)
Table 3-29 Wireless
“Help“ F1 Key Display
NOTE: When both wired and wireless are enabled, the chooses wired.