IRECTION 5535208-100, REV. 2 LOGIQ E9 SERVICE MANUAL
Chapter 8 Replacement procedures 8 - 25
8-4-7 Loading the Application Software Only (cont’d)
NOTE: The window reflects the actual software version instead of “R6.0.6EP1_L9_, as shown above.
4.) As indicated, press any key to continue.
NOTE: When the LoadSoftware Information Window displays asking you to select OK, do nothing. This
window closes automatically.
5.) Wait while the software loads (approximately 7-8 minutes). Do not interrupt this process and ignore
any errors.
After the software load is complete, the system reboots.
Figure 8-15 Command Window
Figure 8-16 LoadSoftware Information Window - Do not select OK
WARNINGWARNING
Do not interrupt the software load at any time.
NOTICE
Do not try to prevent the reboot.

Table of Contents

Other manuals for GE LOGIQ E9

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE LOGIQ E9 and is the answer not in the manual?

GE LOGIQ E9 Specifications

General IconGeneral
TypeUltrasound System
3D/4D ImagingYes
Transducer Ports4 active ports
ConnectivityDICOM, USB, Ethernet
ApplicationsAbdominal, Vascular, Musculoskeletal, Small Parts, Urology, Pediatric, Cardiac, Obstetrics/Gynecology
TransducersLinear, Convex, Phased Array, Volume
Doppler ModesColor Doppler, Power Doppler, Pulsed Wave Doppler, Continuous Wave Doppler
ElastographyYes
Advanced FeaturesB-Flow, Volume Navigation
PortabilityCart-based system
Display19-inch high-resolution LCD monitor

Related product manuals