GE
D
IRECTION 5535208-100, REV. 2 LOGIQ E9 SERVICE MANUAL
8 - 36 Section 8-5 - Replacing Covers and Bumpers
8-5-3 Shearwave Console Cover replacement
NOTE: Whenever the left Side Cover is removed, clear the BEP Fan(s) intake(s) of any debris.
Table 8-7 Manpower / Time and Tools
Manpower /
Time
Tools
One person /
10 minutes
Refer to: 8-2-5 "Tools needed for servicing the LOGIQ E9" on page 8-5.
Table 8-8 Preparations and Preparation Links
Preparations - you must perform the following steps
1. Power down the system.
2. Disconnect the mains power cable from the wall outlet and all Probes and External I/O Cabling.
3. Remove the Left Side Cover and loosen the Top Cover.
Preparation Links (if you need more information):
• 4-2-3 "Power shut down" on page 4-7.
• 8-5-2 "Side Covers replacement" on page 8-30.
• 8-5-4 "Top Cover replacement" on page 8-38.