Loading the software
LOGIQ V2/LOGIQ V1 – Basic Service Manual 7-19
5610739-100 English Rev.10
Loading the System Software with USB memory stick (For R1.0.x)
(continued)
6. Select Finish when the upgrade process completes and the
system will shutdown automatically.
Figure 7-4. System upgrade complete
7. Remove the USB memory stick and then press Power On/
Off switch to power on the system.
NOTE: If you do not remove the USB memory stick, the software
system loading process repeats when the system boots up.