EasyManua.ls Logo

GE PACE Series

GE PACE Series
176 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
PACE Series SCPI Manual
K0472 Revision A 4-79
SOURce
The SOURce subsystem controls the pressure output of the instrument.
SOUR:ALT
Applicability: PACE6000
Command Syntax
SOURce:ALTitude:LEVel:IMMediate:AMPLtude
Parameter: <string>
Short form: SOUR:ALT
Function: Sets the altitude set-point.
Default: -
Query Syntax
SOURce:ALTitude:LEVel:IMMediate:AMPLtude?
Short form: SOUR:ALT?
Function: Queries altitude set-point.
Response: Decimal number representing altitude in current units.
Description
This command sets the altitude set-point. This query gets the decimal number representing
altitude in current units.
Note: If the instrument is not in aero mode it returns an error: -221. Settings conflict.

Table of Contents

Related product manuals