PACE Series SCPI Manual
K0472 Revision A 4-79
SOURce
The SOURce subsystem controls the pressure output of the instrument.
SOUR:ALT
Applicability: PACE6000
Command Syntax
SOURce:ALTitude:LEVel:IMMediate:AMPLtude
Parameter: <string>
Short form: SOUR:ALT
Function: Sets the altitude set-point.
Default: -
Query Syntax
SOURce:ALTitude:LEVel:IMMediate:AMPLtude?
Short form: SOUR:ALT?
Function: Queries altitude set-point.
Response: Decimal number representing altitude in current units.
Description
This command sets the altitude set-point. This query gets the decimal number representing
altitude in current units.
Note: If the instrument is not in aero mode it returns an error: -221. Settings conflict.