EasyManuals Logo

GE PACE Series User Manual

GE PACE Series
176 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #81 background imageLoading...
Page #81 background image
PACE Series SCPI Manual
K0472 Revision A 4-53
:SENS:PRES:INL
Applicability: PACE5000, PACE6000
Command Syntax
n/a
Parameter:
Short form:
Function:
Query Syntax
:SENSe[x][:PRESsure]:INLimits?
where x = 1 or 2 is the module number
(default - 1)
Short form: :SENS:INL?
Function: Query in-limits value
Response: First parameter - current pressure.
Second parameter - in limit:
0 = not in limits
1 = in limits
Description
This command query gets the in-limits value and the in-limits status:
Example
1
TX> :SENS:PRES:INL?
RX> :SENS:PRES:INL 990.0527344, 0
Example
2
TX> :SENS:PRES:INL?
RX> :SENS:PRES:INL 990.0527344, 1