4 Command and Query Summary
K0472 Revision A 4-38
:SENS:ALT:INL:TIME
Applicability: PACE6000
Command Syntax
:SENSe[x][:ALTitude]:INLimits:TIME <number>
where x = 1 or 2 is the module number
(default - 1)
Parameter: <integer> in-limits time in seconds
Short form: :SENS:INL:TIME <number>
Function: Sets the time that the pressure has to be within limits before
generating a service request.
Default: 2 seconds
minimum 2 seconds
maximum 999 seconds
Query Syntax
:SENSe[x][:ALTitude]:INLimits:TIME?
where x = 1 or 2 is the module number
(default - 1)
Short form: :SENS:INL:TIME?
Function: Query in-limits timers.
Response: Number representing in-limits time in seconds.
Description
Sets the in-limits timer value.
Note: If the instrument is not in aero mode it returns an error: -221. Settings conflict.