PACE Series SCPI Manual
K0472 Revision A 4-81
SOUR:MACH:REF:MODE
Applicability: PACE6000
Command Syntax
SOURce:MACH:REFerence:MODE
Parameter: <number>
Short form: SOUR:MACH:REF:MODE
Function: Enables a MACH reference mode to be set.
Default: -
Query Syntax
SOURce:MACH:REFerence:MODE?
Short form: SOUR:MACH:REF:MODE?
Function: Queries the MACH reference mode.
Response: Numeric value representing the MACH reference mode.
Description
This command enables a MACH reference mode to be set. The query gets the MACH
reference entered mode.
Note: If the instrument is not in aero mode it returns an error: -221. Settings conflict.