4 Command and Query Summary
K0472 Revision A 4-84
:SOUR:PRES:COMP
Applicability: PACE5000, PACE6000
Command Syntax
n/a
Parameter:
Short form:
Function:
Default:
Query Syntax
:SOURce[x][:PRESsure]:COMPensate[y]?
where: x (module) = 1 or 2 (default - 1) y (+ve or -ve source pressure) = 1 or 2.
Short form: :SOUR:COMP?
Function: Queries +ve and -ve source pressures.
Response: Pressure value in current pressure units.
Description
This query gets the +ve and -ve source pressures in the current pressure units.
Example
:SOUR:PRES:COMP[x]?
TX> :SOUR:PRES:COMP?
RX> :SOUR:PRES:COMP 3165.9526002
TX> :SOUR:PRES:COMP1?
RX> :SOUR:PRES:COMP 3165.9484591
TX> :SOUR:PRES:COMP2?
RX> :SOUR:PRES:COMP2 -963.9638062
If x is greater than 2, an error will be reported
TX> :SOUR:PRES:COMP3?
TX> :SYST:ERR?
RX> :SYST:ERR -114,"Header suffix out of range"
y=1,2; y Source pressure measurements
1+ve
2-ve