4 Command and Query Summary
K0472 Revision A 4-14
INSTrument
The INSTrument subsystem gets information about the configuration of the instrument .
:INST:CAT:ALL
Applicability: PACE1000, PACE5000, PACE6000
Command Syntax
n/a
Parameter:
Short form:
Function:
Query Syntax
:INSTrument:CATalog[x]:[ALL]?
where: x (module) = 1 or 2 (default - 1)
Short form: :INST:CAT?
Function: Query all ranges fitted
Response: A list of comma separated strings of ranges fitted.
Description
This query returns all the ranges fitted to the instrument. The reply is a comma separated
list of strings representing each range.
Example
for a dual module instrument with a 7.00 barg module and a 3.50barg module both
have a barometric option fitted.
TX> :INST:CAT1?
RX> :INST:CAT: "7.00barg","BAROMETER","8.00bara"
TX> :INST:CAT2?
RX> :INST:CAT: "3.50barg","BAROMETER","4.50bara"